Abstract
Abstract
Objective. The aim of this study was to determine field output correction factors
k
Q
clin
,
Q
ref
f
clin
,
f
ref
and electron fluence perturbation for new PTW unshielded microSilicon and shielded microSilicon X detectors. Approach.
k
Q
clin
,
Q
ref
f
clin
,
f
ref
factors were calculated for 6 and 10 MV with and without flattening filter beams delivered by a TrueBeam STx. Correction factors were determined for field sizes ranging from 0.5 × 0.5 cm2 to 3 × 3 cm2 using both experimental and numerical methods. To better understand the underlying physics of their response, total electron (+positron) fluence spectra were scored in the sensitive volume considering the various component-dependent perturbations. Main results. The microSilicon and microSilicon X detectors can be used down to the smallest studied field size by applying corrections factors fulfilling the tolerance of 5% recommended by the IAEA TRS483. Electron fluence perturbation in both microSilicon detectors was greater than that in water but to a lesser extent than their predecessors. The main contribution of the overall perturbation of the detectors comes from the materials surrounding their sensitive volume, especially the epoxy in the case of unshielded diodes and the shielding for shielded diodes. This work demonstrated that the decrease in the density of the epoxy for the microSilicon led to a decrease in the electron fluence perturbation. Significance. A real improvement was observed regarding the design of the microSilicon and microSilicon X detectors compared to their predecessors.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献