Calorimeter measurements of absolute dose in aluminum, a surrogate of bone, to validate dose-to-medium in Acuros XB

Author:

Szpala StanislawORCID,Renaud JamesORCID,Muir Bryan RORCID,Bourgouin AlexandraORCID,Kohli Kirpal,McEwen MalcolmORCID

Abstract

Abstract Objective. While the accuracy of dose calculations in water with Acuros XB is well established, experimental validation of dose in bone is limited. Acuros XB reports both dose-to-medium and dose-to-water, and these values differ in bone, but there are no reports of measurements of validation in bone. This work compares Acuros XB calculations to measurements of absolute dose in aluminum (medium similar to bone). The validity of using selected relative dosimeters in aluminum is also investigated. Approach. A calorimeter with an aluminum core embedded in an aluminum phantom was selected as bone surrogate for the measurement of absolute dose. Matching the medium of the core to the medium of the phantom allowed eliminating the calculation of the conversion between media. The dose was measured at the fixed depth of 3.3 cm in aluminum (∼9 g·cm−2) with 6X, 10X, 6FFF and 10FFF photon beams from a TrueBeam Varian linac. In addition, experimental cross-calibration between water and aluminum was performed for an IBA CC13 ionization chamber, a PTW microDiamond and EBT3 Gafchromic film. Main results. Calculations with Acuros XB dose-to-medium in aluminum differed from the calorimetry data by −2.8% to −3.5%, depending on the beam. Use of dose-to-water would have resulted in about 39% discrepancy. The cross calibration coefficient between water and aluminum yielded values of about 0.87 for the CC13 chamber, 0.91 for the microDiamond, and 0.88 for the film, and independent of the beam within about ±1%. Significance. It was demonstrated the value of the dose-to-medium in aluminum (surrogate of bone) computed with Acuros XB is close to the value of the absolute dose measured with a calorimeter, and there is a significant discrepancy when dose-to-water is used instead. The use of an ionization chamber, a microDiamond and Gafchromic film in aluminum required a considerable correction from calibration in water.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3