Author:
Chen Yuan,Goorden Marlies C,Beekman Freek J
Abstract
Abstract
SPECT imaging with 123I-FP-CIT is used for diagnosis of neurodegenerative disorders like Parkinson’s disease. Attenuation correction (AC) can be useful for quantitative analysis of 123I-FP-CIT SPECT. Ideally, AC would be performed based on attenuation maps (
μ
-maps) derived from perfectly registered CT scans. Such
μ
-maps, however, are most times not available and possible errors in image registration can induce quantitative inaccuracies in AC corrected SPECT images. Earlier, we showed that a convolutional neural network (CNN) based approach allows to estimate SPECT-aligned
μ
-maps for full brain perfusion imaging using only emission data. Here we investigate the feasibility of similar CNN methods for axially focused 123I-FP-CIT scans. We tested our approach on a high-resolution multi-pinhole prototype clinical SPECT system in a Monte Carlo simulation study. Three CNNs that estimate
μ
-maps in a voxel-wise, patch-wise and image-wise manner were investigated. As the added value of AC on clinical 123I-FP-CIT scans is still debatable, the impact of AC was also reported to check in which cases CNN based AC could be beneficial. AC using the ground truth
μ
-maps (GT-AC) and CNN estimated
μ
-maps (CNN-AC) were compared with the case when no AC was done (No-AC). Results show that the effect of using GT-AC versus CNN-AC or No-AC on striatal shape and symmetry is minimal. Specific binding ratios (SBRs) from localized regions show a deviation from GT-AC
≤
2.5% for all three CNN-ACs while No-AC systematically underestimates SBRs by 13.1%. A strong correlation (
r
≥
0.99) was obtained between GT-AC based SBRs and SBRs from CNN-ACs and No-AC. Absolute quantification (in kBq ml−1) shows a deviation from GT-AC within 2.2% for all three CNN-ACs and of 71.7% for No-AC. To conclude, all three CNNs show comparable performance in accurate
μ
-map estimation and 123I-FP-CIT quantification. CNN-estimated
μ
-map can be a promising substitute for CT-based
μ
-map.
Funder
Netherlands Organization for Scientific Research
Physics Valorization Prize ‘Ultra-fast, ultra-sensitive and ultra-high resolution SPECT’
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献