Real-time marker-less tumor tracking with TOF PET: in silico feasibility study

Author:

Cheng XinyiORCID,Yang Dongxu,Zhong Yuncheng,Shao YipingORCID

Abstract

Abstract Purpose. Although positron emission tomography (PET) can provide a functional image of static tumors for RT guidance, it’s conventionally very challenging for PET to track a moving tumor in real-time with a multiple frame/s sampling rate. In this study, we developed a novel method to enable PET based three-dimension (3D) real-time marker-less tumor tracking (RMTT) and demonstrated its feasibility with a simulation study. Methods. For each line-of-response (LOR) acquired, its positron-electron annihilation position is calculated based on the time difference between the two gamma interactions detected by the TOF PET detectors. The accumulation of these annihilation positions from data acquired within a single sampling frame forms a coarsely measured 3D distribution of positron-emitter radiotracer uptakes of the lung tumor and other organs and tissues (background). With clinically relevant tumor size and sufficient differential radiotracer uptake concentrations between the tumor and background, the high-uptake tumor can be differentiated from the surrounding low-uptake background in the measured distribution of radiotracer uptakes. With a volume-of-interest (VOI) that closely encloses the tumor, the count-weighted centroid of the annihilation positions within the VOI can be calculated as the tumor position. All these data processes can be conducted online. The feasibility of the new method was investigated with a simulated cardiac-torso digital phantom and stationary dual-panel TOF PET detectors to track a 28 mm diameter lung tumor with a 4:1 tumor-to-background 18FDG activity concentration ratio. Results. The initial study shows TOF PET based RMTT can achieve <2.0 mm tumor tracking accuracy with 5 frame s−1 sampling rate under the simulated conditions. In comparison, using reconstructed PET images to track a similar size tumor would require >30 s acquisition time to achieve the same tracking accuracy. Conclusion. With the demonstrated feasibility, the new method may enable TOF PET based RMTT for practical RT applications.

Funder

National Institute of Health

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3