Comprehensive evaluations of a prototype full field-of-view photon counting CT system through phantom studies

Author:

Zhan Xiaohui,Zhang RuoqiaoORCID,Niu Xiaofeng,Hein Ilmar,Budden Brent,Wu Shuoxing,Markov Nicolay,Clarke CameronORCID,Qiang YiORCID,Taguchi Hiroki,Nomura Keiichi,Muramatsu Yoshihisa,Yu Zhou,Kobayashi Tatsushi,Thompson Richard,Miyazaki Hiroaki,Nakai Hiroaki

Abstract

Abstract Objective. Photon counting CT (PCCT) has been a research focus in the last two decades. Recent studies and advancements have demonstrated that systems using semiconductor-based photon counting detectors (PCDs) have the potential to provide better contrast, noise and spatial resolution performance compared to conventional scintillator-based systems. With multi-energy threshold detection, PCD can simultaneously provide the photon energy measurement and enable material decomposition for spectral imaging. In this work, we report a performance evaluation of our first CdZnTe-based prototype full-size PCCT system through various phantom imaging studies. Approach. This prototype system supports a 500 mm scan field-of-view and 10 mm z-coverage at isocenter. Phantom scans were acquired using 120 kVp from 50 to 400 mAs to assess the imaging performance on: CT number accuracy, uniformity, noise, spatial resolution, material differentiation and quantification. Main results. Both qualitative and quantitative evaluations show that PCCT, under the tested conditions, has superior imaging performance with lower noise and improved spatial resolution compared to conventional energy integrating detector (EID)-CT. Using projection domain material decomposition approach with multiple energy bin measurements, PCCT virtual monoenergetic images have lower noise, and good accuracy in quantifying iodine and calcium concentrations. These results lead to increased contrast-to-noise ratio (CNR) for both high and low contrast study objects compared to EID-CT at matched dose and spatial resolution. PCCT can also generate super-high resolution images using much smaller detector pixel size than EID-CT and greatly improve image spatial resolution. Significance. Improved spatial resolution and quantification accuracy with reduced image noise of the PCCT images can potentially lead to better diagnosis at reduced radiation dose compared to conventional EID-CT. Increased CNR achieved by PCCT suggests potential reduction in iodine contrast media load, resulting in better patient safety and reduced cost.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference37 articles.

1. Estimator for photon counting energy selective x-ray imaging with multibin pulse height analysis;Alvarez;Med. Phys.,2011

2. Energy-selective reconstructions in x-ray computerized tomography;Alvarez;Phys. Med. Biol.,1976

3. X-ray detector for spectral photon counting CT;Blevis,2020

4. Photon-counting x-ray detectors for CT;Danielsson;Phys. Med. Biol.,2021

5. A count rate-dependent method for spectral distortion correction in photon counting CT;Dickmann;Proc. SPIE,2018

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3