Abstract
Abstract
The article focuses on the corrosion behavior of medical grade 316L stainless steel used in cardiac stents, orthopaedic, and dental implants. In this research work, electro-discharge treatment (EDT) was used for the deposition of titanium dioxide (TiO2) particles on 316L stainless steel to improve its corrosion-resistant property. The samples were characterized using scanning electron microscopy and x-ray diffraction techniques to correlate the modified surface with the EDT process parameters. The electrochemical corrosion analysis shows a lower corrosion rate for the TiO2 coated specimen (1.883 mpy) compared to the untreated surface (13.729 mpy). This improved corrosion resistance of TiO2 coated substrate was attributed to the formation of various silicides, carbides, and bioactive compounds.
Subject
Materials Chemistry,Surfaces, Coatings and Films,Process Chemistry and Technology,Instrumentation
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献