Abstract
Abstract
Novel marker-based 3D standards, in the form of cascading step-slope pyramids, enable the calibration of three scaling factors and three coupling factors of scanning probe microscopes using only a single sample and a single measurement. They offer better convenience than conventional methods and are therefore increasingly applied. From a quality management perspective, it is of high importance to validate the evaluation algorithm of the 3D calibration strategy and analyze its error contribution. This paper proposes and develops a new software validation method based on both synthetic data and measurement data. Finally, results of our study are discussed, and the impact of the computational error on the measurement uncertainty is examined.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献