Abstract
Abstract
This paper presents an effective method for evaluating the geometric errors of precision translation stages. It quantifies the six degrees of freedom (6-DOF) absolute posture of a moving platform on the stage, where a phase-encoded binary scale is fixed. This evaluation is accomplished using multi-axis readings obtained from three absolute position encoders. The scale errors are compensated using a reversal method, and distances between the measurement points are obtained without the need for additional calibration processes. Therefore, the proposed method can achieve both high accuracy and efficiency, even when implemented for on-site calibration. The performance of the prototype measurement system was evaluated by measuring the 6-DOF geometrics errors of a translation stage over a 100 mm range, and the measurement results showed agreement with the reference data within ±0.3 μm and ±1.2″.
Funder
Length and Dimensional Standards Team
KRISS
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献