Abstract
AbstractObjectiveOur aim was to assess the immediate and short term results of closure of moderate and large patent arterial ducts using the self-expanding and repositionable Amplatzer device.MethodWe attempted closure in 25 patients (10 Females and 15 males) using the Amplatzer occluder. Their median age was 48 months with a range from 8months to 26years and median weight of 14kg with a range from 4.5kg to 48kg. The mean ductal diameter was 4 (S.D 1.51mm). A 6F/7F long sheath was used to deliver the device. Follow up was performed with colour- flow mapping of the pulmonary trunk within 24 hours, at 3 months, and 6 months of closure.ResultsOf the 25 patients, the device was placed successfully in 23. Concurrent angiography showed immediate closure in 12 patients, while 8 had trivial shunting and 3 had mild shunting. Within 24 hours, Doppler examination revealed complete closure in all but three patients, who had a mild residual shunt. Two attempts were unsuccessful. Both these patients underwent successful surgical ligation. All except one patient were discharged on the next day. Of the 23 patients, 15 (65%) have been followed up for 3months, while 8 (35%) have completed 6 months of follow-up. Of the three patients initially with mild residual flow, two had completely closed at 3 months The one remaining patient is yet to be seen at the 3 month follow-up. Thus, at 3 months, all patients studies had shown complete closure.ConclusionAntegrade transcatheter closure using the Amplatzer duct occluder is an efficacious treatment for bigger patent arterial ducts. Long-term follow-up is necessary to show sustained benefits and confirm the absence of side effects.
Publisher
Cambridge University Press (CUP)
Subject
Cardiology and Cardiovascular Medicine,General Medicine,Pediatrics, Perinatology and Child Health
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献