Abstract
The resolution-limiting aberrations of round electromagnetic lenses
can now be successfully overcome via the use of multipole element
“aberration correctors.” The installation and performance of a
hexapole-based corrector (CEOS GmbH) integrated on the probe-forming side
of a JEOL 2200FS FEG STEM/TEM is described. For the resolution of the
microscope not to be severely compromised by its environment, a new,
specially designed building at Oak Ridge National Laboratory has been
built. The Advanced Microscopy Laboratory was designed with the goal of
providing a suitable location for aberration-corrected electron
microscopes. Construction methods and performance of the building are
discussed in the context of the performance of the microscope. Initial
performance of the microscope on relevant specimens and modifications made
to eliminate resolution-limiting conditions are also discussed.
Publisher
Cambridge University Press (CUP)
Reference13 articles.
1. Batson, P.E. (2003).Aberration correction results in the IBM STEM instrument.Ultramicroscopy 96,239–249.
2. Huggins, D.K. , Fellmann, W. , Smith, J.M. , & Kaesz, H.D. (1964).A polynuclear tetracarbonyl hydride of rhenium. Preparation andproperties.J Am Chem Soc 86,4841–4846.
3. Fung, A.S. , Tooley, P.A. , Kelley, M.J. , Koningsberger, D.C. , & Gates, B.C. (1991).Cationic trirhenium rafts on γ-Al2O3:Characterization by X-ray absorption spectroscopy.J Phys Chem 95,225–234.
4. Liu, J. (2004).Advanced electron microscopy characterization of nanostructuredheterogeneous catalysts.Microsc Microanal 10,551–576.
5. Haider, M. , Uhlemann, S. , & Zach, J. (2000).Upper limits for the residual aberrations of a high-resolutionaberration-corrected STEM.Ultramicroscopy 81,163–175.
Cited by
29 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献