Abstract
AbstractThe successful development of third-order aberration correctors in transmission electron microscopy has seen aberration-corrected electron microscopes evolve from specialist projects, custom built at a small number of sites to common instruments in many modern laboratories. Here we describe some initial results illustrating the two- and three-dimensional (3D) performance of an aberration-corrected scanning transmission electron microscope with a prototype improved aberration corrector designed to also minimize fifth-order aberrations and a new, higher brightness gun. We show that atomic columns separated by 0.63 Å can be resolved and demonstrate detection of single dopant atoms with 3D sensitivity.