Ultra-high resolution, 3-dimensional magnetic resonance imaging of the atherosclerotic vessel wall at clinical 7T

Author:

Willemink Martin J.ORCID,Coolen Bram F.ORCID,Dyvorne Hadrien,Robson Philip M.,Bander Ilda,Ishino Seigo,Pruzan Alison,Sridhar Arthi,Zhang Bei,Balchandani Priti,Mani VenkateshORCID,Strijkers Gustav J.ORCID,Nederveen Aart J.,Leiner Tim,Fayad Zahi A.,Mulder Willem J. M.,Calcagno Claudia

Abstract

Accurate quantification and characterization of atherosclerotic plaques with MRI requires high spatial resolution acquisitions with excellent image quality. The intrinsically better signal-to-noise ratio (SNR) at high-field clinical 7T compared to the widely employed lower field strengths of 1.5 and 3T may yield significant improvements to vascular MRI. However, 7T atherosclerosis imaging also presents specific challenges, related to local transmit coils and B1 field inhomogeneities, which may overshadow these theoretical gains. We present the development and evaluation of 3D, black-blood, ultra-high resolution vascular MRI on clinical high-field 7T in comparison lower-field 3T. These protocols were applied for in vivo imaging of atherosclerotic rabbits, which are often used for development, testing, and validation of translatable cardiovascular MR protocols. Eight atherosclerotic New Zealand White rabbits were imaged on clinical 7T and 3T MRI scanners using 3D, isotropic, high (0.63 mm3) and ultra-high (0.43 mm3) spatial resolution, black-blood MR sequences with extensive spatial coverage. Following imaging, rabbits were sacrificed for validation using fluorescence imaging and histology. Image quality parameters such as SNR and contrast-to-noise ratio (CNR), as well as morphological and functional plaque measurements (plaque area and permeability) were evaluated at both field strengths. Using the same or comparable imaging parameters, SNR and CNR were in general higher at 7T compared to 3T, with a median (interquartiles) SNR gain of +40.3 (35.3–80.1)%, and a median CNR gain of +68.1 (38.5–95.2)%. Morphological and functional parameters, such as vessel wall area and permeability, were reliably acquired at 7T and correlated significantly with corresponding, widely validated 3T vessel wall MRI measurements. In conclusion, we successfully developed 3D, black-blood, ultra-high spatial resolution vessel wall MRI protocols on a 7T clinical scanner. 7T imaging was in general superior to 3T with respect to image quality, and comparable in terms of plaque area and permeability measurements.

Funder

American Heart Association

Fulbright Association

Niels Stensen Fellowship

NWO

National Institutes of Health

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Reference57 articles.

1. Heart Disease and Stroke Statistics-2016 Update: A Report From the American Heart Association;M Writing Group;Circulation,2016

2. PET/MR Imaging in Vascular Disease;RS Ripa;Atherosclerosis and Inflammation. PET Clin,2016

3. Computed Tomography and Cardiac Magnetic Resonance in Ischemic Heart Disease;MR Dweck;J Am Coll Cardiol,2016

4. A Review on Atherosclerotic Biology, Wall Stiffness, Physics of Elasticity, and Its Ultrasound-Based Measurement.;AK Patel;Curr Atheroscler Rep,2016

5. Plaque Composition in the Proximal Superficial Femoral Artery and Peripheral Artery Disease Events;MM McDermott;JACC Cardiovasc Imaging,2016

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3