Evaluation of the setup discrepancy between 6D ExacTrac and cone beam computed tomography in spine stereotactic body radiation therapy

Author:

Park Jaehyeon,Yea Ji Woon,Park Jae Won,Oh Se AnORCID

Abstract

The objective of this study was to analyze the difference in residual setup errors between 6D ExacTrac and 3D cone-beam computed tomography (CBCT) image-guided systems in spinal stereotactic body radiation therapy (SBRT). We investigated 76 patients with spinal tumors who received SBRT using Novalis Tx at our institution between January 2013 and September 2020. A Vac-lok (EZ-FIX®, Arlico Medical Company, South Korea) fixture and an assistive device, based on the region involved, were used to immobilize patients and to increase the inter-fractional setup reproducibility. The difference in the root mean square (RMS) between the 6D ExacTrac and 3D CBCT was -0.75 mm, 0.45 mm, 0.16 mm, and -0.03°; the RMS value was 1.31 mm, 1.06 mm, 0.87 mm, and 0.64°; and the standard deviation was 0.80 mm, 0.72 mm, 0.62 mm, and 0.44° for lateral, longitudinal, vertical, and yaw directions, respectively. The difference in the average RMS between ExacTrac and CBCT was <1.03 mm in the translation direction and <0.47° in the rotational direction; the results were statistically significant in the lateral, longitudinal, and vertical directions, but not in the yaw direction. Thus, it is necessary to verify the ExacTrac image according to the CBCT image.

Funder

Yeungnam University

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3