ExacTrac X-Ray 6D Imaging During Stereotactic Body Radiation Therapy of Spinal and Nonspinal Metastases

Author:

Hadj Henni Ahmed1ORCID,Gensanne David1,Bulot Guillaume1,Roge Maximilien1,Mallet Romain1,Colard Elyse1,Daras Marie1,Hanzen Chantal1,Thureau Sebastien1

Affiliation:

1. Centre Henri Becquerel, Rouen, France

Abstract

The objective was to investigate the possibility of using ExacTrac X-ray (ETX) for 6D image guidance in stereotactic body radiation therapy (SBRT) of bone metastasis and to propose a patient management protocol. The analyses were first obtained from measurements on a pelvic phantom and on 19 patients treated for bone metastasis. The phantom study consisted of applying known offsets and evaluating the ETX level of accuracy, where results were compared with kV-cone beam computed tomography (kV-CBCT). Two groups of patients, 10 spinal and 9 nonspinal SBRT cases, were analyzed to evaluate ETX imaging for different bone localisations. A comparison was made between kV-CBCT and ETX prior to the treatment fractions. During treatments, two other kV-CBCT/ETX image pairs were also acquired and a total of 224 shifts were compared. A second study, using the ETX monitoring module, analyzed the intrafraction motion of 8 other patients. In the phantom study, the root mean square (RMS) of the translational and rotational discrepancies between ETX and kV-CBCT were < 0.6 mm and < 0.4°, respectively. For both groups of patients, the RMS of the discrepancies observed between the two imaging systems were greater than the phantom experiment while still remaining < 1 mm and < 0.7°. In the nonspinal group, three patients (2 scapulas and 1 humerus) did not have consistent shift values with ETX due to a lack of anatomical information. When ETX monitoring was used during irradiation, the setup errors measured were on average less than 1 mm/1°. The results obtained validated the use of ETX for 6D image guidance during bone SBRT. Real-time tracking of the target position improves the accuracy of the irradiation. This strategy allowed for faster correction of out-of-tolerance positioning errors. The registration of bone lesions with poor anatomical information is a limitation of this 2D-kV imaging system.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3