Abstract
Closed-system drug transfer devices (CSTDs) are used to prevent occupational exposure to hazardous drugs in health care providers. They are considered Class II medical devices by the US FDA and are cleared but not approved before marketing. While compatibility tests are conducted by CSTD manufacturers, the procuring institution needs to consider performing its own studies before buying these devices. Herein we tested the compatibility of the components of the Needleless® DualGuard CSTD system (vial access clips, vial access spikes, and administration adaptors) with 10 antineoplastic drugs, under simulated clinical conditions, including compounding and administration, and examined drug potency maintenance, plasticizer migration, and device functionality. All drugs maintained potency within 5%. Diisononyl phthalate leakage was observed from the administration adaptors for paclitaxel and concentrated etoposide solution. In addition, white particles were discovered in CSTDs storing busulfan solution and small cracks were observed on devices which stored melphalan. Thus, it was concluded that even in simulated clinical conditions, instead of extreme conditions, there are still concerns regarding the efficacy and safety of CSTD components. The methodology may be used to implement and detect possible interactions between antineoplastic agents and CSTD components before procurement.
Publisher
Public Library of Science (PLoS)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献