Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches

Author:

Jiang YihangORCID,Spies Connor,Magin JustinORCID,Bhosai Satasuk Joy,Snyder Laurie,Dunn JessilynORCID

Abstract

Blood oxygen saturation (SpO2) is an important measurement for monitoring patients with acute and chronic conditions that are associated with low blood oxygen levels. While smartwatches may provide a new method for continuous and unobtrusive SpO2 monitoring, it is necessary to understand their accuracy and limitations to ensure that they are used in a fit-for-purpose manner. To determine whether the accuracy of and ability to take SpO2 measurements from consumer smartwatches is different by device type and/or by skin tone, our study recruited patients aged 18–85 years old, with and without chronic pulmonary disease, who were able to provide informed consent. The mean absolute error (MAE), mean directional error (MDE) and root mean squared error (RMSE) were used to evaluate the accuracy of the smartwatches as compared to a clinical grade pulse oximeter. The percent of data unobtainable due to inability of the smartwatch to record SpO2 (missingness) was used to evaluate the measurability of SpO2 from the smartwatches. Skin tones were quantified based on the Fitzpatrick (FP) scale and Individual Typology Angle (ITA), a continuous measure of skin tone. A total of 49 individuals (18 female) were enrolled and completed the study. Using a clinical-grade pulse oximeter as the reference standard, there were statistically significant differences in accuracy between devices, with Apple Watch Series 7 having measurements closest to the reference standard (MAE = 2.2%, MDE = -0.4%, RMSE = 2.9%) and the Garmin Venu 2s having measurements farthest from the reference standard (MAE = 5.8%, MDE = 5.5%, RMSE = 6.7%). There were also significant differences in measurability across devices, with the highest data presence from the Apple Watch Series 7 (88.9% of attempted measurements were successful) and the highest data missingness from the Withings ScanWatch (only 69.5% of attempted measurements were successful). The MAE, RMSE and missingness did not vary significantly across FP skin tone groups, however, there may be a relationship between FP skin tone and MDE (intercept = 0.04, beta coefficient = 0.47, p = 0.04). No statistically significant difference was found between skin tone as measured by ITA and MAE, MDE, RMSE or missingness.

Funder

AstraZeneca

National Institute of Diabetes and Digestive and Kidney Diseases

Publisher

Public Library of Science (PLoS)

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3