A comparative study of RF heating of deep brain stimulation devices in vertical vs. horizontal MRI systems

Author:

Vu Jasmine,Bhusal Bhumi,Nguyen Bach T.,Sanpitak Pia,Nowac Elizabeth,Pilitsis Julie,Rosenow JoshuaORCID,Golestanirad LalehORCID

Abstract

The majority of studies that assess magnetic resonance imaging (MRI) induced radiofrequency (RF) heating of the tissue when active electronic implants are present have been performed in horizontal, closed-bore MRI systems. Vertical, open-bore MRI systems have a 90° rotated magnet and a fundamentally different RF coil geometry, thus generating a substantially different RF field distribution inside the body. Little is known about the RF heating of elongated implants such as deep brain stimulation (DBS) devices in this class of scanners. Here, we conducted the first large-scale experimental study investigating whether RF heating was significantly different in a 1.2 T vertical field MRI scanner (Oasis, Fujifilm Healthcare) compared to a 1.5 T horizontal field MRI scanner (Aera, Siemens Healthineers). A commercial DBS device mimicking 30 realistic patient-derived lead trajectories extracted from postoperative computed tomography images of patients who underwent DBS surgery at our institution was implanted in a multi-material, anthropomorphic phantom. RF heating around the DBS lead was measured during four minutes of high-SAR RF exposure. Additionally, we performed electromagnetic simulations with leads of various internal structures to examine this effect on RF heating. When controlling for RMS B1+, the temperature increase around the DBS lead-tip was significantly lower in the vertical scanner compared to the horizontal scanner (0.33 ± 0.24°C vs. 4.19 ± 2.29°C). Electromagnetic simulations demonstrated up to a 17-fold reduction in the maximum of 0.1g-averaged SAR in the tissue surrounding the lead-tip in the vertical scanner compared to the horizontal scanner. Results were consistent across leads with straight and helical internal wires. Radiofrequency heating and power deposition around the DBS lead-tip were substantially lower in the 1.2 T vertical scanner compared to the 1.5 T horizontal scanner. Simulations with different lead structures suggest that the results may extend to leads from other manufacturers.

Funder

NIH

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A workflow for predicting radiofrequency‐induced heating around bilateral deep brain stimulation electrodes in MRI;Medical Physics;2023-12-28

2. How Insights from EM Theory Have Guided Surgical Practice to Make MRI Safer for Patients with Active Electronic Implants;2023 International Conference on Electromagnetics in Advanced Applications (ICEAA);2023-10-09

3. Managing Patients With Unlabeled Passive Implants on MR Systems Operating Below 1.5 T;Journal of Magnetic Resonance Imaging;2023-09-28

4. Optimizing the trajectory of deep brain stimulation leads reduces RF heating during MRI at 3 T: Characteristics and clinical translation;2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC);2023-07-24

5. Rapid prediction of MRI-induced RF heating of active implantable medical devices using machine learning;2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC);2023-07-24