Comparison of Bio-Impedance Spectroscopy and Multi-Frequency Bio-Impedance Analysis for the Assessment of Extracellular and Total Body Water in Surgical Patients

Author:

Hannan W. J.1,Cowen S. J.1,Plester C. E.2,Fearon K. C. H.2,Debeau A.2

Affiliation:

1. Department of Medical Physics and Medical Engineering, Western General Hospital, Edinburgh, U.K.

2. University Department of Surgery, Royal Infirmary, Edinburgh, U.K.

Abstract

1. Measurements of extracellular and total body water provide useful information on the nutritional status of surgical patients and may be estimated from whole-body bio-impedance measurements at different frequencies. 2. Resistance and reactance were measured at 50 frequencies from 5 kHz to 1 MHz in 29 surgical patients (17 males, 12 females) with a wide range of extracellular to total body water ratios. 3. A fit to the spectrum of reactance versus resistance data gave predicted resistances at frequencies zero and infinity. Values of extracellular and total body water determined by this bio-impedance spectroscopy technique were regressed against values obtained from radioisotope dilution. The standard errors of the estimate were 1.893l and 3.259l respectively. 4. Resistance indices (height2/resistance) at selected frequencies gave the highest correlations with extracellular and total body water at 5 kHz and 200 kHz respectively, and prediction equations derived from multiple stepwise regressions also showed these to be the optimum frequencies. The standard errors of the estimate for this multi-frequency bio-impedance analysis method were 1.937l and 2.606l for extracellular and total body water respectively. 5. To assess the ability of the two methods to measure changes in extracellular and total body water, reproducibility was assessed from repeat measurements 10 min apart in a subgroup of 15 patients. Bio-impedance spectroscopy gave mean coefficients of variation for extracellular and total body water of 0.9% and 3.0% respectively. For multi-frequency bio-impedance analysis the corresponding coefficients of variation were 0.9% and 0.6%. 6. It is concluded that a simple impedance analyser operating at only two frequencies compares favourably with the more complex spectroscopy technique for the determination of extracellular and total body water in surgical patients.

Publisher

Portland Press Ltd.

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3