Biomaterials for hip implants – important considerations relating to the choice of materials

Author:

Choroszyński Mieczysław,Choroszyński Mieczysław Rorbert,Skrzypek Stanisław Jan

Abstract

AbstractThis article is a review of important material requirements for hip biomaterials including their response to the body environment (biocompatibility), mechanical properties, wear resistance, fretting corrosion and availability as well as the price. The application of proper biomaterials for hip implants is one of the major focal points in this article. Background information is also provided on metals used in other prosthetic devices and implant components. Titanium and its alloys, cobalt base alloys and stainless steels (bio-steels) are used for load-bearing hip implants. These three groups of metallic materials will be introduced and discussed in detail. Metals and their alloys are crystalline materials since their properties depend on the crystal lattice, chemical and phase compositions, grain size, lattice defects, crystalline texture and residual micro- and macro-stresses. All these features of biomaterials are formed during technological manufacturing, such as metallurgical process, solidification, plastic deformation (rolling and forging), machining, heat treatment and coating. All these technological processes work in optimal conditions in order to achieve the optimal microstructure and mechanical, chemical and biological properties. Amongst the above-mentioned particular properties of biomaterials, fretting is a major concern as regards hip implants at the femoral head and neck taper interface. Additional important mechanisms of interaction between the implant and the human body must be taken into account, i.e. diffusion stream of foreign particles and atoms from the implant to body fluids, to the tissue and to the bone. These foreign particles and atoms are released from the implant to the body fluid, to the tissue and to the bone as wear product during use. All together they contribute to the wear, i.e. loss of weight, strength or volume of hip components. Wear rates of ultrahigh molecular weight polyethylene mated against Ti-6Al-4V are significantly greater than the ones for Co-Cr-Mo alloys. Therefore, thermochemical surface treatments like diffusion ion nitriding should be applied to increase the resistance of titanium alloys to wear. Austenitic stainless steels are also used for temporary applications, but they have lower resistance to pitting corrosion than titanium and cobalt alloys. The purpose of the paper is to introduce a group of metallic materials, which is often chosen for surgical hip implants. Conclusions of the paper refer to information which support important medical and patient decisions on hip implants. Also, the development of biomaterials, their treatments, properties, surface layers and coatings are considered. All these features develop over time and need synergy and experience in the progress of the biomedical, mechanical and materials science.

Publisher

Walter de Gruyter GmbH

Subject

Health Informatics,Biochemistry, Genetics and Molecular Biology (miscellaneous),Medicine (miscellaneous),General Computer Science

Reference72 articles.

1. Tech spotlight: biomedical implant alloys;Adv Mater Proc,2001

2. Cast vs. wrought cobalt chromium surgical implant alloys;J Biomed Mater Res,1975

Cited by 27 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3