Author:
Arkani Mohammad,Khalafi Hossein,Vosoughi Naser,Khakshournia Samad
Abstract
AbstractA correlation measuring tool for an endogenous pulsed neutron source experiment is developed in this work. Paroxysmal pulses generated by a bursts of neutron chains are detected by a 10-kbit embedded shift register with a time resolution of 100 ns. The system is implemented on a single reprogrammable device making it a compact, cost-effective instrument, easily adaptable for any case study. The system was verified experimentally in theEsfahan heavy-water zero power reactor(EHWZPR). The results obtained by the measuring tool are validated by the Feynman-α experiment, and a good agreement is seen within the boundaries of statistical uncertainties. The theory of the methods is briefly initiated in the text. Also, the system structure is described, the experimental results and their uncertainties are discussed, and neutron statistics in EHWZPR is examined experimentally.
Reference35 articles.
1. Practical correction of dead time effect in variance - to - mean ratio measurement of;Hazama;Annals Nuclear Energy,2003
2. Renewal Theory;Cox,1962
3. Reference Guide The;Mathworks;Math,2014
4. comparative study of the shift register and conventional coincidence methods for neutron coincidence counting Nuclear Instruments and;Choithramani;Methods,1981
5. Measurement of multiple α - modes at the Delphi subcritical assembly by neutron noise techniques of;Szieberth;Annals Nuclear Energy
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献