Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay
Author:
Bianchi Giorgia1, Frusciante Erika2, Colombo Giulia1ORCID, Infusino Ilenia12, Aloisio Elena1ORCID, Panteghini Mauro12
Affiliation:
1. Clinical Pathology Unit , ASST Fatebenefratelli-Sacco , Milan , Italy 2. Research Centre for Metrological Traceability in Laboratory Medicine (CIRME) , University of Milan , Milan , Italy
Abstract
Abstract
Objectives
Recently, Abbott Diagnostics marketed a new generation of Alinity enzyme assays, introducing a multiparametric calibrator [Consolidated Chemistry Calibrator (ConCC)] in place of or in addition to factor-based calibrations. For alkaline phosphatase (ALP), both calibration options are offered, i.e., with ConCC (ALP2) and with an experimental calibration factor (ALP2F). Both options are declared traceable to the 2011 IFCC reference measurement procedure (RMP). Before to replace the old generation (ALP1) with the new one, we decided to validate the trueness of ALP2/ALP2F.
Methods
Three approaches were employed: (a) preliminary comparison on 48 native frozen serum samples with ALP1, of which traceability to RMP was previously successfully verified; (b) examination of three banked serum pools (BSP) with values assigned by RMP; (c) direct comparison with RMP on a set of 24 fresh serum samples. Bias estimation and regression studies were performed, and the standard measurement uncertainty associated with ALP measurements on clinical samples (uresult) was estimated and compared with established analytical performance specifications (APS). ConCC commutability was also assessed.
Results
A positive proportional bias was found with both ALP2 and ALP2F when compared to ALP1 and RMP. This positive bias was confirmed on BSP: in average, +13.1 % for ALP2 and +10.0 % for ALP2F, respectively. uresult were 13.28 % for ALP2 and 10.04 % for ALP2F, both not fulfilling the minimum APS of 4.0 %. Furthermore, ConCC was not commutable with clinical samples.
Conclusions
Our results unearth problems in the correct implementation of traceability of Alinity ALP2/ALP2F, with the risk for the new assay to be unfit for clinical purposes.
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference36 articles.
1. Panteghini, M. Serum enzymes. In: Rifai, N, Chiu, RWK, Young, I, Burnham, CAD, Wittwer, CT, editors. Tietz textbook of laboratory medicine, 7th ed. St. Louis: Elsevier; 2023:350 p. 2. Infusino, I, Schumann, G, Ceriotti, F, Panteghini, M. Standardization in clinical enzymology: a challenge for the theory of metrological traceability. Clin Chem Lab Med 2010;48:301–7. https://doi.org/10.1515/cclm.2010.075. 3. Infusino, I, Frusciante, E, Braga, F, Panteghini, M. Progress and impact of enzyme measurement standardization. Clin Chem Lab Med 2017;55:334–40. https://doi.org/10.1515/cclm-2016-0661. 4. Schumann, G, Klauke, R, Canalias, F, Bossert-Reuther, S, Franck, PF, Gella, FJ, et al.. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 °C. Part 9: reference procedure for the measurement of catalytic concentration of alkaline phosphatase. Clin Chem Lab Med 2011;49:1439–46. https://doi.org/10.1515/cclm.2011.621. 5. Aloisio, E, Frusciante, E, Pasqualetti, S, Quercioli, M, Panteghini, M. Traceability of alkaline phosphatase measurement may also vary considerably using the same analytical system: the case of Abbott architect. Clin Chem Lab Med 2018;56:135–7. https://doi.org/10.1515/cclm-2017-1007.