Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay

Author:

Bianchi Giorgia1,Frusciante Erika2,Colombo Giulia1ORCID,Infusino Ilenia12,Aloisio Elena1ORCID,Panteghini Mauro12

Affiliation:

1. Clinical Pathology Unit , ASST Fatebenefratelli-Sacco , Milan , Italy

2. Research Centre for Metrological Traceability in Laboratory Medicine (CIRME) , University of Milan , Milan , Italy

Abstract

Abstract Objectives Recently, Abbott Diagnostics marketed a new generation of Alinity enzyme assays, introducing a multiparametric calibrator [Consolidated Chemistry Calibrator (ConCC)] in place of or in addition to factor-based calibrations. For alkaline phosphatase (ALP), both calibration options are offered, i.e., with ConCC (ALP2) and with an experimental calibration factor (ALP2F). Both options are declared traceable to the 2011 IFCC reference measurement procedure (RMP). Before to replace the old generation (ALP1) with the new one, we decided to validate the trueness of ALP2/ALP2F. Methods Three approaches were employed: (a) preliminary comparison on 48 native frozen serum samples with ALP1, of which traceability to RMP was previously successfully verified; (b) examination of three banked serum pools (BSP) with values assigned by RMP; (c) direct comparison with RMP on a set of 24 fresh serum samples. Bias estimation and regression studies were performed, and the standard measurement uncertainty associated with ALP measurements on clinical samples (uresult) was estimated and compared with established analytical performance specifications (APS). ConCC commutability was also assessed. Results A positive proportional bias was found with both ALP2 and ALP2F when compared to ALP1 and RMP. This positive bias was confirmed on BSP: in average, +13.1 % for ALP2 and +10.0 % for ALP2F, respectively. uresult were 13.28 % for ALP2 and 10.04 % for ALP2F, both not fulfilling the minimum APS of 4.0 %. Furthermore, ConCC was not commutable with clinical samples. Conclusions Our results unearth problems in the correct implementation of traceability of Alinity ALP2/ALP2F, with the risk for the new assay to be unfit for clinical purposes.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3