Effect of syringe underfilling on the quality of venous blood gas analysis

Author:

Lippi Giuseppe1ORCID,Pighi Laura1,Tosi Marco1ORCID,Vettori Marco1,Celegon Giovanni1,Favaloro Emmanuel J.234,Salvagno Gian Luca1

Affiliation:

1. Section of Clinical Biochemistry and School of Medicine , University of Verona , Verona , Italy

2. Department of Haematology , Sydney Centres for Thrombosis and Haemostasis, Institute of Clinical Pathology and Medical Research (ICPMR), NSW Health Pathology, Westmead Hospital , Westmead , NSW , Australia

3. School of Medical Sciences, Faculty of Medicine and Health , University of Sydney, Westmead Hospital , Westmead , NSW , Australia

4. Faculty of Science and Health , Charles Sturt University , Wagga Wagga , NSW , Australia

Abstract

Abstract Objectives There is limited information on the influence of collecting small amounts of blood on the quality of blood gas analysis. Therefore, the purpose of this study was to investigate the effects of different degrees of underfilling of syringes on test results of venous blood gas analysis. Methods Venous blood was collected by venipuncture from 19 healthcare workers in three 1.0 mL syringes for blood gas analysis, by manually aspirating different volumes of blood (i.e., 1.0, 0.5 and 0.25 mL). Routine blood gas analysis was then immediately performed with GEM Premier 5,000. The results of the two underfilled syringes were compared with those of the reference syringe filled with appropriate blood volume. Results The values of most assayed parameters did not differ significantly in the two underfilled syringes. Statistically significant variations were found for lactate, hematocrit and total hemoglobin, the values of which gradually increased as the fill volume diminished, as well as for sodium concentration, which decreased in both insufficiently filled blood gas syringes. The bias was clinically meaningful for lactate in syringe filled with 0.25 mL of blood, and for hematocrit, total hemoglobin and sodium in both syringes containing 0.5 and 0.25 mL of blood. Conclusions Collection of smaller volumes of venous blood than the specified filling volume in blood gas syringes may have an effect on the quality of some test results, namely lactate, hematocrit, total hemoglobin and sodium. Specific indications must be given for standardizing the volume of blood to be collected within these syringes.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,Public Health, Environmental and Occupational Health,Health Policy,Medicine (miscellaneous)

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3