Author:
Penders Joris,Fiers Tom,Giri Mimi,Wuyts Birgitte,Ysewyn Larissa,Delanghe Joris R.
Abstract
AbstractBackground: Recently, automated urine test strip readers became available that can report quantitative data. We explored the possibility of measuring all ketone bodies (acetone, acetoacetate, 3-hydroxybutyrate) in urine with these test strips. Monitoring urinary ketone concentrations could offer the advantages of measuring higher values (due to the low renal thresholds) and being less sensitive to fluctuations.Methods: We evaluated URISYS 2400 (Roche) quantitative reflectance data for the ketone reflectance field and compared it with biochemical data from urine samples. Using an easy sample pre-treatment with 3-hydroxybutyrate dehydrogenase, we were able to assay 3-hydroxybutyrate as well, which normally does not react on urine test strips.Results: Within- and between-run reproducibility of the reflectance signal for high- and low-concentration urine pools was 11.0–3.6% and 11.0–5.8% for aceto-acetate, 8.2–9.2% and 10.4–16.1% for acetone, and 5.1–3.0% and 5.6–3.5% for 3-hydroxybutyrate, respectively. The lower limit of detection for acetoacetate was 0.13mmol/L (CV=3.6%). Fair agreement was obtained between test strip data for ketones andcolorimetrically determined acetoacetate values (r=0.90).Conclusions: In urine test strip analysis, quantitative ketone reflectance data allow a simple and fast analysis, offering affordable screening for the detection of ketone body production in diabetes, especially in emergency settings.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献