Affiliation:
1. Department of Diabetes and Endocrinology, Hope Hospital, Salford, U.K.
Abstract
OBJECTIVE: To evaluate the performance of a hand-held ketone sensor that is able to measure blood beta-hydroxybutyrate (beta-HBA) concentrations within 30 s in patients with diabetic ketoacidosis (DKA) and patients who attend a weight management clinic. RESEARCH DESIGN AND METHODS: Two groups of patients were studied: 19 patients admitted with DKA and 156 patients attending a weight management clinic. Paired capillary and venous whole blood samples were measured using the ketone sensor and also using an enzymatic laboratory reference method. RESULTS: The ketone sensor accurately measured beta-HBA concentrations in patients with DKA (limits of agreement -0.9 to + 1.0 mmol/l) or starvation-induced ketonemia (limits of agreement -0.5 to +0.5 mmol/l). CONCLUSIONS: This ketone sensor accurately measures whole blood beta-HBA concentrations within 30 s.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
119 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献