Total haemoglobin – a reference measuring system for improvement of standardisation

Author:

Grote-Koska Denis1,Klauke Rainer12,Kaiser Patricia3,Kramer Udo3,Macdonald Rainer4,Lerche Dietmar5,Staaden Antje12,Brand Korbinian12,Schumann Gerhard12

Affiliation:

1. Medizinische Hochschule Hannover , Institute of Clinical Chemistry , Carl-Neuberg-Str. 1 , 30625 Hannover , Germany

2. Hannover Medical School (MHH) , Hannover , Germany

3. Instand e.V. , Duesseldorf , Germany

4. Physikalisch-Technische Bundesanstalt (PTB) , Berlin , Germany

5. Dr. Lerche KG , Berlin , Germany

Abstract

Abstract Background Total haemoglobin (Hb) concentration in blood belongs to the most requested measurands, and the HiCN method (hemiglobincyanide) is accepted as a reference. Although the reaction principle is clearly characterised, measurement conditions and settings are not consistently defined, some of them influencing the results. An improvement of standardisation is the object. Methods After method optimization, measurement results between different calibration laboratories (CL) were compared with each other and also with results of the National Metrology Institute of Germany (PTB), with target values of certified reference material, within the RELA scheme, and to >1500 results from routine laboratories. Results Overall deviations between three CLs were ≤0.5% (n = 24 samples) in a measurement range of 20 g/L to 300 g/L. A CV of 0.4% was determined in pooled blood (1 year long-term imprecision, 99.0%–101.1% recovery of the mean). For selected measurements (n = 4 samples) the PTB participated without significant differences to three CLs, and no significant differences were observed comparing CLs to certified values of reference materials. The expanded measurement uncertainty (probability 95%) was estimated as 1.1%. Conclusions A reference measuring system, comprising measuring instruments and other devices, including reagents and supply, to generate reference measurement values for total Hb concentration of high accuracy and low measurement uncertainty is presented. Measurement parameters are investigated and defined. The reference measuring system is ready to offer service to EQA providers and to the IVD industry for certifying control materials or calibrators.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine

Reference30 articles.

1. Lewis SM, Garvey B, Manning R, Sharp SA, Wardle J. Lauryl sulphate haemoglobin: a non-hazardous substitute for HiCN in haemoglobinometry. Clin Lab Haemat 1991;13:279–90.

2. Lema OE, Carter JY, Arube PA, Munafu CG, Wangai MW, Rees PH. Evaluation of the alkaline haematin D-575 method for haemoglobin estimation in east Africa. Bull World Health Organ 1994;72:937–41.

3. Moharram NM, El Aouad R, Al Busaidy S, Fabricius A, Heller S, Wood WG, et al. International collaborative assessment study of the AHD575 method for the measurement of blood haemoglobin. East Mediterranean Health J 2006;12:722–34.

4. Shah VB, Shah BS, Puranik GV. Evaluation of non-cyanide methods for haemoglobin estimation. Indian J Pathol Microbiol 2011;54:764–8.

5. Whitehead RD, Mei Z, Mapango C, Jefferds ME. Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings. Ann N Y Acad Sci 2019;1450:147–71.