Jara-Aguirre Jose C.,Baumann Nikola A.,Block Darci R.,Algeciras-Schimnich Alicia
Abstract
Abstract
Background
Heterophile antibody (HAb) interferences in immunoassays can cause falsely elevated hCG concentrations leading to incorrect diagnosis and treatments options. When results are not consistent with the clinical findings, hCG HAb interference investigation may be requested by the physician. A retrospective evaluation of the frequency of HAb interference was performed among cases of physician-requested investigations and the effectiveness of commercially available blocking reagents to detect HAb interference in two immunoassay systems was evaluated.
Methods
One hundred and thirteen physician requests for hCG HAb investigation from 2008 to 2017 were reviewed. The primary method used to measure hCG was the Beckman Coulter Access Total βhCG (2008–2010) and the Roche Elecsys HCG+β (2014–2017). HAb investigation included measurement by two immunoassays before and after treatment of samples with heterophile blocking reagents and serial dilution studies.
Results
Five cases of HAb and HAb-like interference were identified. The interference frequency was 6.7% for the Beckman assay and 2.9% for the Roche assay. The presence of HAb was detected using heterophile blocking reagents and an alternative method in three cases. The other two cases were detected due to discrepant results with an alternative method and non-linear serial dilutions (HAb-like).
Conclusions
HAb interference was observed in the Beckman and the Roche assays. The heterophile blocking reagents failed to detect 40% of interference cases. Blocking reagents should not solely be used for these investigations. Multiple strategies including the use of serial dilutions and using an alternative platform are critical when troubleshooting interferences in hCG immunoassays.
Subject
Biochemistry, medical,Clinical Biochemistry,General Medicine