Author:
Liu Jiakai,Tan Chin Hon,Badrick Tony,Loh Tze Ping
Abstract
AbstractBackground:Recently, the total prostate-specific antigen (PSA) assay used in a laboratory had a positive bias of 0.03 μg/L, which went undetected. Consequently, a number of post-prostatectomy patients with previously undetectable PSA concentrations (defined as <0.03 μg/L in that laboratory) were being reported as having detectable PSA, which suggested poorer prognosis according to clinical guidelines.Methods:Through numerical simulations, we explored (1) how a small bias may evade the detection of routine quality control (QC) procedures with specific reference to the concentration of the QC material, (2) whether the use of ‘average of normals’ approach may detect such a small bias, and (3) describe the use of moving sum of number of patient results with detectable PSA as an adjunct QC procedure.Results:The lowest QC level (0.86 μg/L) available from a commercial kit had poor probability (<10%) of a bias of 0.03 μg/L regardless of QC rule (i.e. 1:2S, 2:2S, 1:3S, 4:1S) used. The average number of patient results affected before error detection (ANPed) was high when using the average of normals approach due to the relatively wide control limits. By contrast, the ANPed was significantly lower for the moving sum of number of patient results with a detectable PSA approach.Conclusions:Laboratory practitioners should ensure their QC strategy can detect small but critical bias, and may require supplementation of ultra-low QC levels that are not covered by commercial kits with in-house preparations. The use of moving sum of number of patient results with a detectable result is a helpful adjunct QC tool.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Cited by
38 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献