Failure of Current Laboratory Protocols to Detect Lot-to-Lot Reagent Differences: Findings and Possible Solutions

Author:

Algeciras-Schimnich Alicia1,Bruns David E2,Boyd James C2,Bryant Sandra C3,La Fortune Kristin A2,Grebe Stefan KG1

Affiliation:

1. Division of Clinical Biochemistry and Immunology, Department of Laboratory, Medicine and Pathology, and

2. Division of Clinical Pathology, Department of Pathology, University of Virginia Health System, Charlottesville, VA

3. Division of Biomedical Statistics and Informatics, Department of Health Sciences Research, Mayo Clinic, Rochester, MN

Abstract

BACKGROUND Maintaining consistency of results over time is a challenge in laboratory medicine. Lot-to-lot reagent changes are a major threat to consistency of results. METHODS For the period October 2007 through July 2012, we reviewed lot validation data for each new lot of insulin-like growth factor 1 (IGF-1) reagents (Siemens Healthcare Diagnostics) at Mayo Clinic, Rochester, MN, and the University of Virginia, Charlottesville, VA. Analyses of discarded patient samples were used for comparison of lots. For the same period, we determined the distributions of reported patient results for each lot of reagents at the 2 institutions. RESULTS Lot-to-lot validation studies identified no reagent lot as significantly different from the preceding lot. By contrast, significant lot-to-lot changes were seen in the means and medians of 105 668 reported patient IGF-I results during the period. The frequency of increased results increased nearly 2-fold to a high of 17%, without detectable changes in the underlying patient demographics. Retrospective statistical analysis indicated that lot-to-lot comparison protocols were underpowered and that validation studies for this assay required testing >100 samples to achieve 90% power to detect reagent lots that would significantly alter the distributions of patient results. CONCLUSIONS The number of test samples required for adequate lot-to-lot validation protocols is high and may be prohibitively large, especially for low-volume or complex assays. Monitoring of the distributions of patient results has the potential to detect lot-to-lot inconsistencies relatively quickly. We recommend that manufacturers implement remote monitoring of patient results from analyzers in multiple institutions to allow rapid identification of between-lot result inconsistency.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

Cited by 91 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3