Establishing normal values of total testosterone in adult healthy men by the use of four immunometric methods and liquid chromatography-mass spectrometry

Author:

Montagna GiacomoORCID,Balestra Samuela,D’Aurizio Federica,Romanelli Francesco,Benagli Cinzia,Tozzoli Renato,Risch Lorenz,Giovanella Luca,Imperiali Mauro

Abstract

Abstract Background: The total testosterone (T) cutoffs clinically adopted to define late-onset hypogonadism (LOH) do not consider the differences that exist between different analytical platforms, nor do they consider the body mass index (BMI) or age of the patient. We aimed at providing method, age and BMI-specific normal values for total T in European healthy men. Methods: A total of 351 eugonadal healthy men were recruited, and total T was measured with four automated immunometric assays (IMAs): ARCHITECT i1000SR (Abbott), UniCel DxI800 (Beckman Coulter), Cobas e601 (Roche), IMMULITE 2000 (Siemens) and liquid chromatography-tandem mass spectrometry (LC-MS/MS). Reference ranges (RRs) were calculated for each method. Results: Passing and Bablok regression analysis and Bland-Altman plot showed an acceptable agreement between Abbott and LC-MS/MS, but a poor one between LC-MS/MS and the other IMAs. Age-specific T concentrations in non-obese (BMI <29.9 kg/m2) men were greater than in all men. The total T normal range, in non-obese men aged 18–39 years, measured with LC-MS/MS was 9.038–41.310 nmol/L. RRs calculated with LC-MS/MS statistically differed from the ones calculated with all individual IMAs, except Abbott and among all IMAs. Statistically significant differences for both upper and lower reference limits between our RRs and the ones provided by the manufacturers were also noticed. Conclusions: We calculated normal ranges in a non-obese cohort of European men, aged 18–39 years, with four commercially available IMAs and LC-MS/MS and found statistically significant differences according to the analytical method used. Method-specific reference values can increase the accuracy of LOH diagnosis and should be standardly used.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Cited by 15 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3