Design and analysis of lattice structure applied humerus semi-prosthesis

Author:

Savran Efe1,Kalay Onur Can1,Alp Nazmi Bülent2,Karpat Fatih1

Affiliation:

1. Mechanical Engineering , Bursa Uludag University , Bursa , 16059 , Türkiye

2. Orthopedics & Traumatology , Bursa City Hospital , Bursa , Türkiye

Abstract

Abstract Bone tissue loss may occur in bone structures, which are one of the elements that provide the body’s endurance and movement of living things, due to situations such as falling, hitting, or cancer formation. In bad scenarios, applications such as an external plate or internal rod addition are made to regain the old durability of the structure. At the same time, full or semi-prosthesis applications can be made in cases where the original bone structure cannot be preserved. With today’s advanced possibilities, lattice structures can be produced effortlessly with the additive manufacturing (AM) method. Here, the formation of the structure that can show anisotropic behavior depending on the production and the effect of the roughness caused by the production quality should also be seen in the process plan. In this study, it was aimed to compare the durability of titanium (Ti-6Al-4V) and magnesium (ZK60) materials for humeral half prosthesis using cubic-based lattice structure and to show their differences compared to the original bone structure. Maximum stress and deformation values were obtained by performing analyses with the finite element method on the lattice semi-humerus prosthesis obtained with this aim. Reliability analysis was made on the data obtained, and parameter optimization of the lattice structure was aimed. As a result of the study, it was seen that the lattice structure with 65% porosity compared to the reference values is reliable and with the same reliability rate, magnesium provides approximately 60% lightness compared to titanium.

Funder

The Scientific and Technological Research Council of Turkey

Yükseköğretim Kurulu

Publisher

Walter de Gruyter GmbH

Subject

Mechanical Engineering,Mechanics of Materials,General Materials Science

Reference75 articles.

1. M. Peto, O. Aguilar-Rosas, E. Ramirez-Cedillo, M. Jimenez, A. Hernandez, and H. R. Siller, “A proof of concept study of the mechanical behavior of lattice structures used to design a shoulder hemi-prosthesis,” J. Eng. Sci. Med. Diagn. Ther., vol. 4, no. 3, pp. 1–11, 2021, https://doi.org/10.1115/1.4051419.

2. K. Willemsen, T. D. Berendes, T. Geurkink, et al.., “A novel treatment for anterior shoulder instability: a biomechanical comparison between a patient-specific implant and the latarjet procedure,” J. Bone Jt. Surg. Am. Vol., vol. 101, no. 14, pp. 1–8, 2019, https://doi.org/10.2106/JBJS.18.00892.

3. H. Sofu, S. Gürsu, N. Koçkara, A. Öner, A. Issın, and Y. Çamurcu, “Recurrent anterior shoulder instability: review of the literature and current concepts,” World J. Clin. Cases, vol. 2, no. 11, p. 676, 2014, https://doi.org/10.12998/wjcc.v2.i11.676.

4. J. Corona-Castuera, D. Rodriguez-Delgado, J. Henao, J. C. Castro-Sandoval, and C. A. Poblano-Salas, “Design and fabrication of a customized partial hip prosthesis employing CT-scan data and lattice porous structures,” ACS Omega, vol. 6, pp. 6902–6913, 2021, https://doi.org/10.1021/acsomega.0c06144.

5. P. Agaba, B. J. Kildow, H. Dhotar, T. M. Seyler, and M. Bolognesi, “Comparison of postoperative complications after total hip arthroplasty among patients receiving aspirin, enoxaparin, warfarin, and factor Xa inhibitors,” J. Orthop., vol. 14, no. 4, pp. 537–543, 2017, https://doi.org/10.1016/j.jor.2017.08.002.