Abstract
Abstract
Background
Reprocessing of flexible endoscopes (FEs) is often expensive, time consuming, and becomes increasingly complex, due to rising demands of hygiene. After beneficial results in reprocessing of rigid endoscopes using Impelux™ UV-C light technology, we tested the same method for reprocessing of FEs without working channel.
Materials and methods
Testing was performed on FEs without working channel after routine clinical use (transnasal flexible endoscopy). Disinfection consisted of mechanical precleaning and 60 s exposure to Impelux™ UV-C light technology. Bacterial contamination was tested on 50 FEs before and after disinfection. Further 50 FEs regarding protein residuals. The absolute effectiveness of the D60 was tested on 50 test bodies (RAMS) with a standardized contamination of 107 colony-forming units (CFU) of Enterococcus faecium.
Results
The FEs were contaminated with a high average value of 916.7 CFU (± 1057 CFU) after clinical usage. After reprocessing, an average contamination of 2.8 CFU (± 1.6) on 14% (n = 7) of the FEs was detected consisting of non-pathogenic species, the remaining FE were sterile. After reprocessing, all FEs were protein-free (< 1 μg). The artificially contaminated test bodies showed no remaining bacterial contamination after disinfection, resulting in an average absolute germ reduction of about 107 CFU.
Conclusion
Impelux™ UV-C light technology efficiently reduces bacterial contamination of FEs and might be useful in daily practice.
Funder
Philipps-Universität Marburg
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Otorhinolaryngology
Reference24 articles.
1. Rudhart SA, Günther F, Dapper L, Thangavelu K, Gehrt F, Stankovic P, Wilhelm T, Guenzel T, Stuck BA, Hoch S (2020) UV light-based decontamination: an effective and fast way for disinfection of endoscopes in otorhinolaryngology? Eur Arch Otorhinolaryngol 277(8):2363–2369. https://doi.org/10.1007/s00405-020-05978-w
2. Seoane-Vazquez E, Rodriguez-Monguio R (2008) Endoscopy-related infection: relic of the past? CurrOpin Infect Dis 21(4):362–366. https://doi.org/10.1097/QCO.0b013e328301396b
3. Kenters N, Huijskens EG, Meier C, Voss A (2015) Infectious diseases linked to cross-contamination of flexible endoscopes. Endosc Int Open 3(4):E259-265. https://doi.org/10.1055/s-0034-1392099
4. Spaulding EH (1968) Chemical disinfection of medical and surgical materials. In: Lawrence CBS (ed) Disinfection, Sterilization, and Preservation. Lea & Febiger, Philadelphia, pp 517–531
5. Lerner J (2007) Ensuring the effective reprocessing of flexible endoscopes. Health Dev 36(11):352–361