Abstract
Abstract
Background
A gamma probe is a handheld device used for intraoperative interventions following interstitial injection of a radiotracer to locate regional lymph nodes through the external detection of radiation. This work reports on the design and performance evaluation of a novel fully integrated gamma probe (GammaPen), recently developed by our group.
Materials and methods
GammaPen is an all-in-one pocket gamma probe with low weight and adequate dimensions, consisting of a detector, a control unit and output all together. The detector module consists of a cylindrical Thallium-activated Cesium Iodide [CsI (Tl)] crystal optically coupled to a Silicon photomultiplier (SiPM), shielded using Tungsten housing on side and back faces. The electronics of the probe consists of two small boards to handle signal processing and analog peak detection tasks. A number of parameters, including probe sensitivity in air/water, spatial resolution in air/water, angular resolution in air/water, and side and back shielding effectiveness, were measured to evaluate the performance of the probe based on NEMA NU3-2004 standards.
Results
The sensitivity of the probe in air at distances of 10, 30, and 50 mm is 18784, 3500, and 1575 cps/MBq. The sensitivity in scattering medium was also measured at distances of 10, 30, and 50 mm as 17,680, 3050, and 1104 cps/MBq. The spatial and angular resolutions in scattering medium were 47 mm and 87 degree at 30 mm distance from the probe, while they were 40 mm and 77 degree in air. The detector shielding effectiveness and leakage sensitivity are 99.91% and 0.09%, respectively.
Conclusion
The performance characterization showed that GammaPen can be used effectively for sentinel lymph node localization. The probe was successfully used in several surgical interventions by an experienced surgeon confirming its suitability in a clinical setting.
Funder
Private Foundation of Geneva University Hospitals
Université de Genève
Publisher
Springer Science and Business Media LLC
Subject
Radiology Nuclear Medicine and imaging,General Medicine
Reference32 articles.
1. Gershenwald JE, Thompson W, Mansfield PF, Lee JE, Colome MI, Tseng Ch, et al. Multi-institutional melanoma lymphatic mapping experience: the prognostic value of sentinel lymph node status in 612 stage I or II melanoma patients. J Clin Oncol. 1999;17:976.
2. Stoeckli SJ, Steinert H, Pfaltz M, Schmid S. Sentinel lymph node evaluation in squamous cell carcinoma of the head and neck. Otolaryngol Head Neck Surg. 2001;125:221–6.
3. Veronesi U, Paganelli G, Galimberti V, Viale G, Zurrida S, Bedoni M, et al. Sentinel-node biopsy to avoid axillary dissection in breast cancer with clinically negative lymph-nodes. Lancet. 1997;349:1864–7.
4. Bullock T (2001) Surgical treatment—evidence-based and problem-oriented: Edited by Rene G. Holzheimer, John A. Mannick, München, Germany: W. Zuckschwerdt Verlag Munchen. 843 pages. Price $99.95, 2002, Elsevier.
5. Jewell EL, et al. Detection of sentinel lymph nodes in minimally invasive surgery using indocyanine green and near-infrared fluorescence imaging for uterine and cervical malignancies. Gynecol Oncol. 2014;133(2):274–7.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献