Abstract
Abstract
Introduction
Lower transcatheter aortic valve replacement (TAVR) pressure gradients have been reported after implantation of self-expanding valves compared with balloon-expandable valves; however, there is a paucity of data on the relationship between invasively measured transvalvular pressure gradients and Doppler-derived measurements.
Methods
From September 2013 to September 2018, patients with native aortic valve stenosis who had both intraoperative invasive and postoperative echocardiography transvalvular pressure gradients were included for analysis. We used parametric and nonparametric statistics to compare aortic gradients within and between groups.
Results
Of 171 patients, 152 (88.9%) patients had TAVR with a balloon-expandable valve and 19 (11.1%) with a self-expanding valve. Among all patients, the invasive aortic gradient was 7.8 ± 3.2 mmHg and the Doppler-derived aortic gradient was 11.0 ± 4.5 mmHg (p < 0.001). Among those who received a balloon-expandable valve, the invasive aortic gradient was 7.5 ± 3 mmHg and the Doppler aortic gradient was 11.4 ± 4.5 mmHg (p < 0.001). In contrast, among patients who received a self-expanding valve, the invasive aortic gradient was 10.3 ± 3.4 mmHg and the Doppler aortic gradient was 8.5 ± 4.6 mmHg (p = 0.18).
Conclusions
Balloon-expandable valves were associated with lower invasive measurements versus post-TAVR Doppler gradients, while results were inconclusive regarding self-expanding valves.