Abstract
Abstract
Background
Device-related thrombus (DRT) after left atrial appendage occlusion (LAAO) is potentially linked to adverse events. Although clinical reports suggest an effect of the device type and position on the DRT risk, in-depth studies of its mechanistic basis are needed. This in silico study aimed to assess the impact of the position of non-pacifier (Watchman) and pacifier (Amulet) LAAO devices on surrogate markers of DRT risk.
Methods
The LAAO devices were modeled with precise geometry and virtually implanted in different positions into a patient-specific left atrium. Using computational fluid dynamics, the following values were quantified: residual blood, wall shear stress (WSS) and endothelial cell activation potential (ECAP).
Results
In comparison to an ostium-fitted device position, deep implantation led to more residual blood, lower average WSS and higher ECAP surrounding the device, especially on the device’s atrial surface and the surrounding tissue, suggesting increased risk for potential thrombus. For the non-pacifier device, an off-axis device orientation resulted in even more residual blood, higher ECAP and similar average WSS as compared to an ostium-fitted device position. Overall, the pacifier device showed less residual blood, higher average WSS and lower ECAP, compared to the non-pacifier device.
Conclusions
In this in silico study, both LAAO device type and implant position showed an impact on potential markers of DRT in terms of blood stasis, platelet adhesion and endothelial dysfunction. Our results present a mechanistic basis for clinically observed risk factors of DRT and the proposed in silico model may aid in the optimization of device development and procedural aspects.
Graphical abstract
Funder
Universitätsklinikum Bonn
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,General Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献