Abstract
Abstract
Purpose
Extracorporeal membrane oxygenation has gained increasing attention in the treatment of patients with acute and chronic cardiopulmonary and respiratory failure. However, clotting within the oxygenators or other components of the extracorporeal circuit remains a major complication that necessitates at least a device exchange and bears risks of adverse events for the patients. In order to better predict thrombus growth within oxygenators, we present an approach for in-vitro visualization of thrombus growth using real-time X-ray imaging.
Methods
An in-vitro test setup was developed using low-dose anticoagulated ovine blood and allowing for thrombus growth within 4 h. The setup was installed in a custom-made X-ray setup that uses phase-contrast for imaging, thus providing enhanced soft-tissue contrast, which improves the differentiation between blood and potential thrombus growth. During experimentation, blood samples were drawn for the analysis of blood count, activated partial thromboplastin time and activated clotting time. Additionally, pressure and flow data was monitored and a full 360° X-ray scan was performed every 15 min.
Results
Thrombus formation indicated by a pressure drop and changing blood parameters was monitored in all three test devices. Red and white thrombi (higher/lower attenuation, respectively) were successfully segmented in one set of X-ray images.
Conclusion
We showed the feasibility of a new in-vitro method for real-time thrombus growth visualization by means of phase contrast X-ray imaging. In addition, with more blood parameters that are clinically relevant, this approach might contribute to improved oxygenator exchange protocols in the clinical routine.
Funder
START Program Medical Faculty RWTH Aachen University
RWTH Aachen University
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Biomedical Engineering
Reference32 articles.
1. Antoine, E., C. Buchanan, K. Fezzaa, W.-K. Lee, M. N. Rylander, and P. Vlachos. Flow measurements in a blood-perfused collagen vessel using x-ray micro-particle image velocimetry. PLoS ONE. 2013. https://doi.org/10.1371/journal.pone.0081198.
2. Arens, J., M. Schoberer, A. Lohr, T. S. M. Orlikowsky, R. K. Jellema, J. J. Collins, et al. NeonatOx: a pumpless extracorporeal lung support for premature neonates. Artif. Organs. 35:997–1001, 2011. https://doi.org/10.1111/j.1525-1594.2011.01324.x.
3. Biological evaluation of medical devices—Selection of tests for interactions with blood: Biological evaluation of medical devices—Part 4: Selection of tests for interactions with blood 04/2017;11.100.20. 2017th ed. Berlin: DIN Deutsches Institut für Normung e. V.
4. Blok, S. L. J., G. E. Engels, and W. van Oeveren. In vitro hemocompatibility testing: the importance of fresh blood. Biointerphases. 11:29802, 2016. https://doi.org/10.1116/1.4941850.
5. Borggrefe, J., J. Kottlors, M. Mirza, V.-F. Neuhaus, N. Abdullayev, V. Maus, et al. Differentiation of clot composition using conventional and dual-energy computed tomography. Clin. Neuroradiol. 28:515–522, 2018. https://doi.org/10.1007/s00062-017-0599-3.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献