Abstract
Abstract
Purpose
In vitro stimulation of native and bioartificial vessels in perfusable systems simulating natural mechanical environments of the human vasculature represents an emerging approach in cardiovascular research. Promising results have been achieved for applications in both regenerative medicine and etiopathogenetic investigations. However, accurate and reliable simulation of the wide variety of physiological and pathological pressure environments observed in different vessels still remains an unmet challenge.
Methods
We established a modular hemodynamic simulator (MHS) with interchangeable and modifiable components suitable for the perfusion of native porcine—(i.e. the aorta, brachial and radial arteries and the inferior vena cava) and bioartificial fibrin-based vessels with anatomical site specific pressure curves. Additionally, different pathological pressure waveforms associated with cardiovascular diseases including hyper- and hypotension, tachy- and bradycardia, aortic valve stenosis and insufficiency, heart failure, obstructive cardiomyopathy and arterial stiffening were simulated. Pressure curves, cyclic distension and shear stress were measured for each vessel and compared to ideal clinical pressure waveforms.
Results
The pressure waveforms obtained in the MHS showed high similarity to the ideal anatomical site specific pressure curves of different vessel types. Moreover, the system facilitated accurate emulation of physiological and different pathological pressure conditions in small diameter fibrin-based vessels.
Conclusion
The MHS serves as a variable in vitro platform for accurate emulation of physiological and pathological pressure environments in biological probes. Potential applications of the system include bioartificial vessel maturation in cardiovascular tissue engineering approaches as well as etiopathogenetic investigations of various cardiovascular pathologies.
Funder
German Society for Implant Research and Development
Medizinische Hochschule Hannover (MHH)
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Biomedical Engineering
Reference43 articles.
1. Aper, T., M. Wilhelmi, U. Boer, S. Lau, N. Benecke, A. Hilfiker, et al. Dehydration improves biomechanical strength of bioartificial vascular graft material and allows its long-term storage. Innov. Surg. Sci. 3(3):215–224, 2018.
2. Bergh, N., M. Ekman, E. Ulfhammer, M. Andersson, L. Karlsson, and S. Jern. A new biomechanical perfusion system for ex vivo study of small biological intact vessels. Ann. Biomed. Eng. 33(12):1808–1818, 2005.
3. Chambers, D. Venous pressure waveforms. In: Basic Physiology for Anaesthetists, edited by D. Chambers, C. Huang, and G. Matthews. Cambridge: Cambridge University Press, 2015, pp. 161–163.
4. Chouinard, J. A., S. Gagnon, M. G. Couture, A. Levesque, and P. Vermette. Design and validation of a pulsatile perfusion bioreactor for 3D high cell density cultures. Biotechnol. Bioeng. 104(6):1215–1223, 2009.
5. Denardo, S. J., R. Nandyala, G. L. Freeman, G. L. Pierce, and W. W. Nichols. Pulse wave analysis of the aortic pressure waveform in severe left ventricular systolic dysfunction. Circ. Heart Fail. 3(1):149–156, 2010.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献