Abstract
Abstract
Purpose
To evaluate the influence of cortical and cancellous bone structure on the biomechanical properties of all-suture and conventional anchors and compare the morphological bone damage after their failure. The hypothesis of the study is that all-suture anchor pullout is less invasive and that the pullout force is influenced by the cortical thickness.
Methods
Thirty human humeri were biomechanically tested as follows: starting with a load cycle from 20 to 50 N, a stepwise increase of the upper peak force by 0.05 N for each cycle at a rate of 1 Hz was performed. Analysis included maximum pullout strength for three different anchor implantation angles (45°, 90°, 110°) of the two anchor types. After anchor pullout, every sample underwent micro-CT analysis. Bone mineral density (BMD) and cortical thickness were determined at the anchor implantation site. Furthermore, the diameter of the cortical defect and the volume of the bone cavity were identified.
Results
The maximum pullout strength of all-suture anchors demonstrates a strong correlation to the adjacent cortical thickness (r = 0.82, p ≤ 0.05) with at least 0.4 mm needed to withstand 200 N. No correlation could be seen in conventional anchors. Moreover, no correlation could be detected for local BMD in both anchors. All-suture anchors show a significantly narrower cortical defect as well as a smaller bone cavity following pullout (4.3 ± 1.3 mm vs. 5.3 ± 0.9 mm, p = 0.037; 141 mm3 vs. 212 mm3; p = 0.009). The cortical defect is largest if the anchors are placed at a 45° angle.
Conclusion
In contrast to conventional anchors, the pullout force of all-suture anchors depends on the thickness of the humeral cortex. Furthermore, all-suture anchors show a significantly smaller cortical defect as well as decreased bone damage in the case of pullout. Therefore, the clinical implication of this study is that all-suture anchors are advantageous due to their bone preserving ability. Also, intraoperative decortication should not be performed and cortical thickness should be preoperatively evaluated to decrease the risk of anchor failure.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
28 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献