Author:
Reiff Florian Sebastian,Bischoff Charlotte,Woelfler Henriette,Roehling Stefan
Abstract
Abstract
Purpose
To investigate whether inexperienced users applying a static navigation system can perform in-vitro a fully guided implant placement protocol and achieve similar results in terms of accuracy compared to experienced clinicians.
Methods
Based on 36 identical resin models, a computer-assisted implant planning was performed and a surgical guide was produced accordingly. Three study groups were composed with 12 operators, each: control group with experienced surgeons (DOC), test group 1 with dental technicians (TEC) and test group 2 with non-specialists (OFC). Using a fully guided drilling protocol, two implants were placed into each of the 36 models. Subsequently, the differences between the virtually planned and final implant positions were determined and the transfer accuracy was evaluated.
Results
For the control group DOC, the mean value of axial deviation was 1.90 ± 1.15 degrees, for 3-dimensional deviation at the implant base 0.52 ± 0.33 mm, for 3-dimensional deviation at the implant tip 0.76 ± 0.39 mm and for vertical deviation at the implant tip − 0.11 ± 0.51 mm. For corresponding parameters, the mean values of test group TEC were 1.99 ± 0.87 degrees, 0.42 ± 0.21 mm, 0.68 ± 0.30 mm and − 0.03 ± 0.33 mm and for test group OFC 2.29 ± 1.17 degrees, 0.63 ± 0.35 mm, 0.89 ± 0.43 mm and − 0.24 ± 0.57 mm, respectively. The results did not reveal any statistically significant differences between the control and the 2 test groups (p˃0.05).
Conclusion
The results of the present in-vitro study demonstrated that inexperienced users applying a static navigation system can perform a fully guided implant placement protocol and achieve similar results in terms of accuracy compared to experienced clinicians in this specific in vitro setup.
Funder
Universitätsmedizin Greifswald
Publisher
Springer Science and Business Media LLC
Reference37 articles.
1. Bover-Ramos F, Viña-Almunia J, Cervera-Ballester J, Peñarrocha-Diago M, García-Mira B (2018) Accuracy of Implant Placement with Computer-guided surgery: a systematic review and Meta-analysis comparing cadaver, clinical, and in Vitro studies. Int J Oral Maxillofac Implants 33(1):101–115
2. Gargallo-Albiol J, Barootchi S, Marqués-Guasch J, Wang HL (2020) Fully guided Versus Half-guided and Freehand Implant Placement: systematic review and Meta-analysis. Int J Oral Maxillofac Implants 35(6):1159–1169
3. Gargallo-Albiol J, Barootchi S, Salomó-Coll O, Wang HL (2019) Advantages and disadvantages of implant navigation surgery. A systematic review. Annals Anat = Anatomischer Anzeiger: Official Organ Anatomische Gesellschaft 225:1–10
4. Schnutenhaus S, Brunken L, Edelmann C, Dreyhaupt J, Rudolph H, Luthardt RG (2020) Alveolar ridge preservation and primary stability as influencing factors on the transfer accuracy of static guided implant placement: a prospective clinical trial. BMC Oral Health 20(1):178
5. Schelbert T, Gander T, Blumer M, Jung R, Rücker M, Rostetter C (2019) Accuracy of computer-guided template-based Implant surgery: a computed tomography-based clinical Follow-Up study. Implant Dent 28(6):556–563