Abstract
OBJECTIVE—Hypoglycemia is a common acute complication of diabetes therapy. The GlucoWatch biographer provides frequent and automatic glucose measurements with an adjustable low-glucose alarm. We have analyzed the performance of the biographer low-glucose alarm relative to hypoglycemia as defined by blood glucose ≤3.9 mmol/l.
RESEARCH DESIGN AND METHODS—The analysis was based on 1,091 biographer uses from four clinical trials, which generated 14,487 paired (biographer and blood glucose) readings.
RESULTS—The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose ≤3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose >3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer.
CONCLUSIONS—The frequent and automatic nature of the biographer readings allows more effective detection of hypoglycemia than that achieved with current medical practice.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
104 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献