Abstract
OBJECTIVE—To compare pre- and postmeal capillary blood glucose concentrations measured at the finger, forearm, and thigh in adults with diabetes.
RESEARCH DESIGN AND METHODS—For phase 1, capillary blood glucose concentrations were measured at six time points (premeal and at ∼60, 90, 120, 150, and 180 min postmeal) using a blood glucose monitoring system and technician-obtained samples collected from finger, forearm, and thigh sites of 42 adults with diabetes. The finger samples were also tested with a laboratory instrument. For phase 2, ∼14 weeks later, the testing procedures were repeated with 38 subjects from the original study population.
RESULTS—Meter finger results were accurate at all time points. Alternate sites tended to produce lower glucose readings compared to finger readings at times when glucose was increasing rapidly (60 and 90 min postmeal). Forearm-to-finger differences correlated with rates of glucose change (r = 0.56, P < 0.001), as did the thigh-to-finger differences (r = 0.52, P < 0.001). Other factors, such as subject age, BMI, diabetes type, and insulin dependence did not have a significant impact on site differences. When the testing procedures were repeated with the same subjects, the pattern of site differences was consistent, although individual results were variable.
CONCLUSIONS—Changes in blood glucose immediately after a meal may be identified at finger sites before detection at forearm or thigh sites. Alternate site testing appears to be a useful option for routine self-monitoring before meals; however, patients and clinicians should recognize that results may be different from fingertip results when glucose levels are changing rapidly.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Reference9 articles.
1. Koschinsky T, Heidkamp P, Vogt H: FreeStyle: technical evaluation of a new blood glucose meter for self-monitoring of blood glucose (Abstract). Diabetes 49(Suppl. 1):A114, 2000
2. Louie RF, Tang Z, Sutton DV, Lee JH, Kost GJ: Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med 124:257–266, 2000
3. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL: Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 10:622–628, 1987
4. Larsson-Cohn U: Differences between capillary and venous blood glucose during oral glucose tolerance tests. Scand J Clin Lab Invest 36:805–808, 1976
5. Eriksson KF, Fex G, Trell E: Capillary-venous differences in blood glucose values during the oral glucose tolerance test. Clin Chem 29:993, 1983
Cited by
119 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献