Point-of-Care Glucose Testing

Author:

Louie Richard F.1,Tang Zuping1,Sutton Demetria V.1,Lee Judith H.1,Kost Gerald J.1

Affiliation:

1. From the Departments of Medical Pathology (Mr Louie, Dr Tang, Mss Sutton and Lee, and Dr Kost) and Clinical Chemistry (Ms Lee and Dr Kost), School of Medicine, University of California, Davis, Calif.

Abstract

Abstract Objective.—To assess the clinical performance of glucose meter systems when used with critically ill patients. Design.—Two glucose meter systems (SureStepPro and Precision G) and a modular adaptation (Immediate Response Mobile Analysis-SureStepPro) were assessed clinically using arterial samples from critically ill patients. A biosensor-based analyzer (YSI 2700) and a hospital chemistry analyzer (Synchron CX-7) were the primary and secondary reference instruments, respectively. Patients and Setting.—Two hundred forty-seven critical care patients at the University of California, Davis, Medical Center participated in this study. Outcome Measures.—Error tolerances of ±15 mg/dL for glucose levels ≤100 mg/dL and ±15% for glucose levels >100 mg/dL were used to evaluate glucose meter performance; 95% of glucose meter measurements should fall within these tolerances. Results.—Compared to the primary reference method, 98% to 100% of SureStepPro and 91% to 95% of Precision G measurements fell within the error tolerances. Paired differences of glucose measurements versus critical care variables (Po2, pH, Pco2, and hematocrit) were analyzed to determine the effects of these variables on meter measurements. Po2 and Pco2 decreased Precision G and SureStepPro measurements, respectively, but not enough to be clinically significant based on the error tolerance criteria. Hematocrit levels affected glucose measurements on both meter systems. Modular adaptation did not affect test strip performance. Conclusions.—Glucose meter measurements correlated best with primary reference instrument measurements. Overall, both glucose meter systems showed acceptable performance for point-of-care testing. However, the effects of some critical care variables, especially low and high hematocrit values, could cause overestimated or underestimated glucose measurements.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3