Affiliation:
1. Department of Pediatrics, Medical College of Wisconsin, Milwaukee Children's Hospital 1700 West Wisconsin Avenue, Milwaukee, Wisconsin 53233
Abstract
Twenty-four insulin-dependent diabetic pediatric subjects were studied for 1444 nights for detection of nocturnal hypoglycemia with the Teledyne Sleep Sentry (Teledyne Avionics, Charlottesville, Virginia): a wristwatch-like unit that measures absolute changes in skin temperature and decreases in galvanic skin resistance, indicators of hypoglycemia. The device detected 42 of 46 recognized hypoglycemic episodes. One hundred fifty alarms were sounded without evidence of hypoglycemia, probably due to night sweating. Twenty-five percent of the subjects experienced unacceptable cutaneous reactions, presumably due to metallic iontophoresis.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
14 articles.