Comparison of a Needle-Type and a Microdialysis Continuous Glucose Monitor in Type 1 Diabetic Patients

Author:

Wentholt Iris M.1,Vollebregt Marit A.1,Hart Augustus A.2,Hoekstra Joost B.1,DeVries J. Hans1

Affiliation:

1. Department of Internal Medicine, Academic Medical Center, Amsterdam, the Netherlands

2. Department of Clinical Epidemiology and Biostatistics, Academic Medical Center, Amsterdam, the Netherlands

Abstract

OBJECTIVE—We examined the reliability of two continuous glucose sensors in type 1 diabetic patients at night and during rapid glucose excursions and verified the hypothesized nocturnal hypoglycemic drift of the needle-type sensor (CGMSgold) and delay of the microdialysis sensor (GlucoDay). RESEARCH DESIGN AND METHODS—Blood was sampled overnight twice per hour in 13 patients. Rapid-acting insulin was given subcutaneously 30 min after breakfast. Sampling once per minute started 45 min after breakfast and 75 min after insulin injection for 30 min, with the aim of determining peak and nadir glucose values. Mean absolute differences (MADs) between sensor and blood glucose values were calculated. Sensor curves were modeled for all patients using linear regression. Horizontal and vertical shifts of sensor curves from the blood glucose curves were assessed. A vertical shift indicates sensor drift and a horizontal shift sensor delay. RESULTS—Drift was minimal in the needle-type and microdialysis sensors (−0.02 and −0.04 mmol/l). Mean ± SD delay was 7.1 ± 5.5 min for the microdialysis sensor (P < 0.001). MAD was 15.0% for the needle-type sensor and 13.6% for the microdialysis sensor (P = 0.013). After correction for the 7-min delay, the microdialysis MAD improved to 11.7% (P < 0.0001). CONCLUSIONS—The microdialysis sensor was more accurate than the needle-type sensor, with or without correction for a 7-min delay. In contrast to the previous version, the current needle-type sensor did not exhibit nocturnal hypoglycemic drift. Continuous subcutaneous glucose sensors are valuable adjunctive tools for glucose trend analyses. However, considering the large MADs, individual sensor values should be interpreted with caution.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 99 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3