Clinical Performance of CGMS in Type 1 Diabetic Patients Treated by Continuous Subcutaneous Insulin Infusion Using Insulin Analogs

Author:

Guerci Bruno1,Floriot Michèle1,Böhme Philip1,Durain Danielle1,Benichou Muriel1,Jellimann Stéphanie1,Drouin Pierre1

Affiliation:

1. Service de Diabétologie, Maladies Métaboliques et Maladies de la Nutrition, Centre d’Investigation Clinique-INSERM/CHU de NANCY-Hôpital Jeanne d’Arc, Centre Hospitalo-Universitaire de Nancy, Toul cedex, France

Abstract

OBJECTIVE—Performance criteria have been established for in vitro blood glucose monitoring, particularly for the self-monitoring of blood glucose using glucose meters. Devices intended for use in the future, such as the continuous glucose monitoring system (CGMS), should satisfy similar criteria, particularly in diabetic patients under intensive therapy. RESEARCH DESIGN AND METHODS—The analysis was conducted on 18 type 1 diabetic patients (not controlled, HbA1c >7.5%) treated by external pump using insulin analogs. Each patient received a glucose sensor for 3 days during his/her hospitalization and was instructed in its operation. Medtronic criteria were used to determine the accuracy of the CGMS. In addition, the data were analyzed according to American Diabetes Association (ADA) criteria, Clarke Error Grid analysis, and method of residuals, with the glucose oxidase method using a Beckman analyzer used as the reference method. Specificity and sensitivity were evaluated from the viewpoint of accuracy in the detection of hypoglycemia. For nine patients, two glucose sensors were simultaneously inserted into an abdominal site to determine the reproducibility of the system. RESULTS—Among the 33 glucose sensors inserted, 6 (18%) were nonfunctional. The mean duration of CGMS recording was 63 ± 12 h. From all of the 692 sets of data that paired glucose readings and CGMS, the coefficients of correlation ranged from 0.87 to 0.92 and the mean absolute error ranged from 12.8 to 15.7%. The time experienced in hypoglycemia (<55 mg/dl) was reported at 86 ± 62 min/day. Only 39% of the CGMS values satisfied the ADA precision criteria to within ±10%, and 19% of these values satisfied the future ADA precision criteria of accuracy to within ±5%. The means of difference method showed that the CGMS slightly underestimated the plasma glucose values (mean = −12 mg/dl). Error grid analysis showed only 77% of the glucose sensor values were in zone A, and 98.9% were in zones A and B. Two values fell in zone C and a single value fell in zone D. The sensitivity and specificity of the CGMS to detect hypoglycemia were 33 and 96%, respectively. A total of 6,666 paired sensor values were recorded with a coefficient of correlation of 0.84 with a coefficient of variation of 8.25%. CONCLUSIONS—CGMS could be useful in routine clinical practice to provide much more information on the glucose profile than intermittent self-monitoring of blood glucose (SMBG). However, CGMS cannot be used as a replacement for glucose meters because it does not satisfy the conventional performance goals set down for in vitro glucose measurements and could therefore lead to clinically incorrect treatment decisions.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3