Total Hip Wear Assessment: A Comparison Between Computational and In Vitro Wear Assessment Techniques Using ISO 14242 Loading and Kinematics

Author:

Matsoukas George1,Willing Ryan1,Kim Il Yong1

Affiliation:

1. Department of Mechanical and Materials Engineering, Queen’s University, McLaughlin Hall 305, 130 Stuart Street, Kingston, ON, K7L 3N6, Canada

Abstract

In the present study a direct comparison was made between in vitro total hip wear testing and a computational analysis considering the effects of time and a nonlinear stress-strain relationship for ultrahigh molecular weight polyethylene (UHMWPE) at 37°C. The computational simulation was made correct through calibration to experimental volumetric wear results, and the predicted damage layout on the acetabular liner surface was compared with results estimated from laser scanning of the actual worn specimens. The wear rates for the testing specimens were found to be 17.14±1.23 mg/106 cycles and 19.39±0.79 mg/106 cycles, and the cumulative volumetric wear values after 3×106 cycles were 63.70 mm3 and 64.02 mm3 for specimens 1 and 2, respectively. The value of the calibrated wear coefficient was found to be 5.32(10−10) mm3/N mm for both specimens. The major difference between the computational and experimental wear results was the existence of two damage vectors in the experimental case. The actual location of damage was virtually the same in both cases, and the maximum damage depth of the computational model agreed well with the experiment. The existence of multiple wear vectors may indicate the need for computational approaches to account for multidirectional sliding or strain hardening of UHMWPE. Despite the limitation in terms of describing the overall damage layout, the present computational model shows that simulation can mimic some of the behavior of in vitro wear.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference39 articles.

1. Hip Implants: Paper V. Physiological Effects;Goswami;Mater. Des.

2. Biological Response to Wear Debris: Cellular Interaction Causing Osteolysis;Shanbhag

3. Effects of Implant Design Parameters on Fluid Convection, Potentiating Third-Body Debris Ingress Into the Bearing Surfaces During THA Impingement/Subluxation;Lundberg;J. Biomech.

4. Aseptic Loosening;Wooley;Gene Ther.

5. The Problem is Osteolysis;Harris;Clin. Orthop. Relat. Res.

Cited by 27 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3