Experimental Validation of a Finite Element Model of the Proximal Femur Using Digital Image Correlation and a Composite Bone Model

Author:

Dickinson A. S.1,Taylor A. C.2,Ozturk H.1,Browne M.1

Affiliation:

1. Bioengineering Research Group, University of Southampton, Highfield, Southampton, Hants SO17 1BJ, UK

2. Finsbury Development Limited, 13 Mole Business Park, Randalls Road, Leatherhead, Surrey KT22 0BA, UK

Abstract

Computational biomechanical models are useful tools for supporting orthopedic implant design and surgical decision making, but because they are a simplification of the clinical scenario they must be carefully validated to ensure that they are still representative. The goal of this study was to assess the validity of the generation process of a structural finite element model of the proximal femur employing the digital image correlation (DIC) strain measurement technique. A finite element analysis model of the proximal femur subjected to gait loading was generated from a CT scan of an analog composite femur, and its predicted mechanical behavior was compared with an experimental model. Whereas previous studies have employed strain gauging to obtain discreet point data for validation, in this study DIC was used for full field quantified comparison of the predicted and experimentally measured strains. The strain predicted by the computational model was in good agreement with experimental measurements, with R2 correlation values from 0.83 to 0.92 between the simulation and the tests. The sensitivity and repeatability of the strain measurements were comparable to or better than values reported in the literature for other DIC tests on tissue specimens. The experimental-model correlation was in the same range as values obtained from strain gauging, but the DIC technique produced more detailed, full field data and is potentially easier to use. As such, the findings supported the validity of the model generation process, giving greater confidence in the model’s predictions, and digital image correlation was demonstrated as a useful tool for the validation of biomechanical models.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Cited by 78 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3