Characterizing the HeartMate II Left Ventricular Assist Device Outflow Using Particle Image Velocimetry

Author:

Rowlands Grant W.1,Good Bryan C.1,Deutsch Steven1,Manning Keefe B.23

Affiliation:

1. Department of Biomedical Engineering, The Pennsylvania State University, 205 Hallowell Building, University Park, PA 16802

2. Professor Department of Biomedical Engineering, The Pennsylvania State University, 205 Hallowell Building, University Park, PA 16802;

3. Department of Surgery, Penn State Hershey Medical Center, Hershey, PA 17033 e-mail:

Abstract

Ventricular assist devices (VADs) are implanted in patients with a diseased ventricle to maintain peripheral perfusion as a bridge-to-transplant or as destination therapy. However, some patients with continuous flow VADs (e.g., HeartMate II (HMII)) have experienced gastrointestinal (GI) bleeding, in part caused by the proteolytic cleavage or mechanical destruction of von Willebrand factor (vWF), a clotting glycoprotein. in vitro studies were performed to measure the flow located within the HMII outlet cannula under both steady and physiological conditions using particle image velocimetry (PIV). Under steady flow, a mock flow loop was used with the HMII producing a flow rate of 3.2 L/min. The physiological experiment included a pulsatile pump operated at 105 BPM with a ventricle filling volume of 50 mL and in conjunction with the HMII producing a total flow rate of 5.0 L/min. Velocity fields, Reynolds normal stresses (RNSs), and Reynolds shear stresses (RSSs) were analyzed to quantify the outlet flow's potential contribution to vWF degradation. Under both flow conditions, the HMII generated principal Reynolds stresses that are, at times, orders of magnitude higher than those needed to unfurl vWF, potentially impacting its physiological function. Under steady flow, principal RNSs were calculated to be approximately 500 Pa in the outlet cannula. Elevated Reynolds stresses were observed throughout every phase of the cardiac cycle under physiological flow with principal RNSs approaching 1500 Pa during peak systole. Prolonged exposure to these conditions may lead to acquired von Willebrand syndrome (AvWS), which is accompanied by uncontrollable bleeding episodes.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference30 articles.

1. Heart Disease and Stroke Statistics—At-a-Glance;Am. Heart Assoc.,2015

2. OPTN/SRTR 2015 Annual Data Report: Heart;Am. J. Transplant.,2017

3. Ventricular Assist Devices;Kirk,2014

4. Science for Surgeons: Understanding Pump Thrombogenesis in Continuous-Flow Left Ventricular Assist Devices;J. Thorac. Cardiovasc. Surg.,2015

5. Acquired von Willebrand Syndrome in Patients With a Centrifugal or Axial Continuous Flow Left Ventricular Assist Device;JACC Heart Failure,2014

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3