Computational Fluid Dynamics (CFD) Study of the 4th Generation Prototype of a Continuous Flow Ventricular Assist Device (VAD)
Author:
Song Xinwei1, Wood Houston G.1, Olsen Don2
Affiliation:
1. Department of Mechanical and Aerospace Engineering, University of Virginia, Charlottesville, Virginia USA 2. Utah Artificial Heart Institute, Salt Lake City, Utah USA
Abstract
The continuous flow ventricular assist device (VAD) is a miniature centrifugal pump, fully suspended by magnetic bearings, which is being developed for implantation in humans. The CF4 model is the first actual prototype of the final design product. The overall performances of blood flow in CF4 have been simulated using computational fluid dynamics (CFD) software: CFX, which is commercially available from ANSYS Inc. The flow regions modeled in CF4 include the inlet elbow, the five-blade impeller, the clearance gap below the impeller, and the exit volute. According to different needs from patients, a wide range of flow rates and revolutions per minute (RPM) have been studied. The flow rate-pressure curves are given. The streamlines in the flow field are drawn to detect stagnation points and vortices that could lead to thrombosis. The stress is calculated in the fluid field to estimate potential hemolysis. The stress is elevated to the decreased size of the blood flow paths through the smaller pump, but is still within the safe range. The thermal study on the pump, the blood and the surrounding tissue shows the temperature rise due to magnetoelectric heat sources and thermal dissipation is insignificant. CFD simulation proved valuable to demonstrate and to improve the performance of fluid flow in the design of a small size pump.
Publisher
ASME International
Subject
Physiology (medical),Biomedical Engineering
Reference28 articles.
1. American Medical Association, 2002, “Heart Disease and Stroke Statistics—2003 Update,” American Heart Association, Dallas, TX. 2. DeBakey, M. E.
, 2000, “The Odyssey of the Artificial Heart,” Artif. Organs, 24(6), pp. 405–11. 3. Olsen, D. B.
, 2000, “The History of Continuous-Flow Blood Pumps,” Artif. Organs, 24(6), pp. 401–4. 4. Song, X., Throekmorton, A. L., Untaroiu, A., Patel, S., Allaire, P. E., Wood, H. G., and Olsen, D. B., 2003, “Axial Flow Blood Pumps,” ASAIO J., 49, pp. 355–364. 5. Burgreen, G. W., Antaki, J. F., and Griffith, B. P., 1996, “A Design Improvement Strategy for Axial Blood Pumps Using Computational Fluid Dynamics,” ASAIO J., 42, pp. M354–M360M354–M360.
Cited by