Inductive Sensing to Detect Tissue Thickness Between Magnets for Potential Application in Magnetic Compression Based Anastomosis

Author:

Arun Anupama12,Gaston Brandon3,Kwiat Dillon3,Imamura-Ching Jill3,Fechter Richard3,Chen Shijie4,Jiang Hao4,Harrison Michael3,Roy Shuvo5

Affiliation:

1. Department of Surgery, University of California San Francisco, San Francisco, CA 94143;

2. Department of Bioengineering and Therapeutic Sciences, University of California San Francisco, San Francisco, CA 94143 e-mail:

3. Department of Surgery, University of California San Francisco, San Francisco, CA 94143

4. School of Electrical Engineering, San Francisco State University, San Francisco, CA 94132

5. Department of Bioengineering and Therapeutic Sciences, University of California San Francisco, San Francisco, CA 94143

Abstract

Magnetic compression based anastomoses use magnetic force to necrose tissue between two magnets to create an anastomosis. Nickel-plated neodymium–iron–boron magnets are used in our study. The compression pressure between the magnets depends on the distance between the magnets, which is determined by the thickness of the compressed tissue and depends on bowel wall thickness and elasticity. It is critical to know the distance between the magnets once the tissue is compressed because the magnets must be within a critical distance of each other in order to create enough compressive force to necrose the tissue. We have developed an inductance sensor to detect the distance (tissue thickness) between the two magnets after the surgeon has deployed them. Inductance sensing is a contact-less sensing method that enables precise short-range detection of conducting surfaces. The inductor coil mounted on one magnet detects the second magnet by measuring the change in inductance due to eddy current induced on the nickel-plated surface of the second magnet. The change in the inductance is proportional to the change in distance between the magnets. The sensor was first calibrated by using polycarbonate sheets to simulate the intestine tissue. We are able to detect up to 6 mm of spacing between the magnets. Pig intestine from Yorkshire pigs was used to characterize the sensor. We are able to distinguish up to five distinct layers of the intestine from the large intestine. This sensing mechanism can indicate the operating surgeon the exact thickness of the tissue compressed between the two magnets. The surgeon can thus be sure of formation of a clean anastomosis and avoid the likelihood of the magnets sliding away or uncoupling.

Publisher

ASME International

Subject

Biomedical Engineering,Medicine (miscellaneous)

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3