Calibration and Application of an Intra-articular Force Transducer for the Measurement of Patellar Tendon Graft Forces: An in Situ Evaluation

Author:

Fleming B. C.1,Good L.2,Peura G. D.1,Beynnon B. D.1

Affiliation:

1. McClure Musculoskeletal Research Center, Department of Orthopædics and Rehabilitation, University of Vermont, Burlington, VT 05405

2. Sports Trauma Research Group, Department of Orthopædics, University Hospital, Linko¨ping, Sweden

Abstract

The objective of this study was to evaluate two calibration methods for the “Arthroscopically Implantable Force Probe” (AIFP) that are potentially suitable for in vivo use: (1) a direct, experimentally based method performed by applying a tensile load directly to the graft after it is harvested but prior to implantation (the “pre-implantation” technique), and (2) an indirect method that utilizes cadaver-based analytical expressions to transform the AIFP output versus anterior shear load relationship, which may be established in vivo, to resultant graft load (the “post-implantation” technique). The AIFP outputs during anterior shear loading of the knee joint using these two calibration methods were compared directly to graft force measurements using a ligament cutting protocol and a 6 DOF load cell. The mean percent error ((actual – measured)/(actual) * 100) associated with the pre-implantation calibration ranged between 85 and 175 percent, and was dependent on the knee flexion angle tested. The percent error associated with the post-implantation technique was evaluated in two load ranges: loads less than 40 N, and loads greater than 40 N. For graft force values greater than 40 N, the mean percent errors inherent to the post-implantation calibration method ranged between 20 and 29 percent, depending on the knee flexion angle tested. Below 40 N, these errors were substantially greater. Of the two calibration methods evaluated, the post-implantation approach provided a better estimate of the ACL graft force than the pre-implantation technique. However, the errors for the post-implantation approach were still high and suggested that caution should be employed when using implantable force probes for in vivo measurement of ACL graft forces.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3